AF46 UV Analyzer

Dual Wavelength UV Absorption Analyzer

The AF46 is a dual wavelength, high precision UV light absorption analyzer designed for inline operation to provide accurate concentration measurements with remarkable repeatability, linearity and resolution. Wavelength and optical path length can be individually configured to specific requirements for the highest accuracy.

Used For

  • UV range:
    • 0 - 0.05 to 2 CU (concentration units), dependent on filter/OPL used
  • UV absorption control and UV Absorption control
    • in Chromatography: Pre / Post column monitoring (OD) for optimized pooling
    • of concentrations in permeate or retentate
    • in TFF / Ultrafiltration
    • of organic load in water / condensate
    • of the concentration of aromatics (Caprolactam, Aniline, Phenol, etc.)
    • as TOC, COD, BOD trend analysis
  • Disinfection control treatment of water
Features
Benefits

UV analyzer for dual wavelength absorbance measurement with discrete lamp references
Simultaneous precise UV measurements at 2 wavelengths with compensation of lamp effects

Various UV wavelengths available
Optimized solution for specific applications requirements

Prepared for use of calibration filters
Easy validation by user with NIST traceable calibrated filters

Long lasting light source
reliable process monitoring with reduced maintenance

High temperature version available
Suitable for SIP process

The AF46 is a dual wavelength UV analyzer ideally suited to monitor protein concentrations (e.g., monoclonal antibodies) in downstream biotechnology applications such as chromatography or filtration. Using two different wavelengths the sensor is capable of detecting very low protein concentrations for accurate pooling and at the same time is able to monitor the whole elution peak with very high optical density (OD) using a second wavelength.

Besides protein measurements the sensor is often used to monitor nucleic acid contaminations of protein solutions. Due to the modularity different optical path lengths (OPL) and wavelengths can be combined to measure any product absorbing in the UV range.

AF46
Schematic
  1. Sensor Body
  2. Windows
  3. Beam splitter
  4. Filter A
  5. Measurement detector A
  6. Filter B
  7. Measurement detector B
  8. Reference detector A
  9. Reference detector B
  10. Lamp module (mercury)
Measurement
Measurement principle

2- Channel Absorption of light

Measurement wavelength(s)

254-13 / 280-13, 254-13 / 313-13, 280-09 / 300-05, 280-09 / 300-13, 280-09 / 313-13, 280-13 / 300-13, 280-13 / 313-13, 290-13 / 313-13 nm, ... others on request

Detector(s)

2 silicon photodiodes (hermetically sealed)

Reference detector(s)

2 silicon photodiodes (hermetically sealed)

Measuring Range

any measuring range between

0 - 0.05 to 2 CU

(dependent on filter used / OPL)

(contact product specialist for application specific ranges)

Optical path length

1 - 160 mm

Calibration

CU (concentration units) application specific calibration

Light source

low pressure mercury lamp

typical life span: 1 to 2 years (8,000 to 16,000 hours)

Resolution

< ± 0.05 % of respective measuring range

Repeatability

< ± 0.5 % of respective measuring range

Linearity

< ± 1 % of respective measuring range (specific to application)

Protection

all optical parts have an IP rating of IP65 or higher

Sensor Body
Material

Stainless steel 1.4435 (SS 316L), 1.4539 (904 L), 1.4571 (SS 316Ti), 1.4462 (318 LN), Titanium 3.7035 (Grade 2), Hastelloy 2.4602 (C22), Plastic TFM™ 4215, ... others on request

Line size

1/8 in. to 6 in. (DN 6 to DN 150), ... others on request

Process connection

Flanges (ASME, DIN, EN, JIS), Clamps (TC, ISO, DIN), Female Threads (NPT, DIN), Sanitary Threads (DIN 11851), Tube Ends (DIN, ISO, OD),DIN 11864-1/-2/-3 (DIN, ISO, OD), Varivent (DIN, IPS, OD), ... others on request

Process pressure

0 - 100 bar (0 - 1450 psi) - higher on request depending on process connection, materials and design

Windows

2-Sapphire, 3-Sapphire Biotech (do not use Pyrex®)

Window gaskets

Viton® (FDA), EPDM / Kalrez® 6230 (FDA / USP Class VI), Kalrez® 4079, ... others on request (do not use Silicone)

Temperature Ratings
Process temperature

permanent: 0 - 70 °C ( 32 - 158 °F)

peak 15 min/day: 0 - 135 °C ( 32 - 275 °F)

peak 30 min/day: 0 - 120 °C ( 32 - 248 °F)

Process temperature
OPTION HT

permanent: -30 - 120 °C (-22 - 248 °F)

peak 15 min/day: -30 - 150 °C (-22 - 302 °F)

peak 30 min/day: -30 - 140 °C (-22 - 284 °F)

Process temperature
OPTION EX

permanent: -30 - 70 °C (-22 - 158 °F)

peak 15 min/day: -30 - 135 °C (-22 - 275 °F)

peak 30 min/day: -30 - 120 °C (-22 - 248 °F)

Process temperature
OPTION EX-HT

permanent: -30 - 120 °C (-22 - 248 °F)

peak 15 min/day: -30 - 150 °C (-22 - 302 °F)

peak 30 min/day: -30 - 140 °C (-22 - 284 °F)

Ambient temperature

permanent: 0 - 40 °C ( 32 - 104 °F)

operation: -30 - 40 °C (-22 - 104 °F) with options HT / EX / EX-HT

transport: -20 - 70 °C ( -4 - 158 °F)

Explosion Proof
Ex-proof

none

Ex-proof
OPTION EX (EN-D)

Sensor assembly in ex-proof version accor. ATEX (EN-D)

Ex-proof
OPTION EX (FM-D)

Sensor assembly in ex-proof version accor. FM (FM-D)

Calibration
Calibration adapter VB

Filter adapter FH03 for calibration filter used for sensor verification

Pressure and temperature ratings specified herein may be subject to limitations - see instruction manual. The appropriate choice of material for all wetted parts is the sole responsibility of the user. Data given are subject to changes without prior notice.

Speed / Spare Parts

Lamp Module AF45/46     PN: 1426-3139-1604-02

Lamp Module AF45/46

PN: 1426-3139-1604-02

Installation Kit AF/TF PN: 1201-3131-0004-00

Installation Kit AF/TF

PN: 1201-3131-0004-00

O-Ring 10.10 x 1.60 Viton® (4) PN: 1203-0004-0008-02

O-Ring 10.10 x 1.60 Viton® (4)

PN: 1203-0004-0008-02

O-Ring 18.77 x 1.78 Viton® (4) PN: 1203-0004-0013-02

O-Ring 18.77 x 1.78 Viton® (4)

PN: 1203-0004-0013-02

O-Ring 21.95 x 1.78 Viton® (FDA) (4)  PN: 1203-0004-0015-02

O-Ring 21.95 x 1.78 Viton® (FDA) (4)

PN: 1203-0004-0015-02

O-Ring 25.12 x 1.78 Viton® (4)  PN: 1203-0004-0016-02

O-Ring 25.12 x 1.78 Viton® (4)

PN: 1203-0004-0016-02

O-Ring 31.47 x 1.78 Viton® (4)  PN: 1203-0004-0018-02

O-Ring 31.47 x 1.78 Viton® (4)

PN: 1203-0004-0018-02

O-Ring 50.52 x 1.78 Viton® (4) PN: 1203-0004-0021-02

O-Ring 50.52 x 1.78 Viton® (4)

PN: 1203-0004-0021-02

Product Variants

AF46-HT-VB
AF46-HT-VB
AF46-EX-HT-VB
AF46-EX-HT-VB

Process temperature OPTION HT

permanent: -30 - 120 °C (-22 - 248 °F)
peak 15 min/day: -30 - 150 °C (-22 - 302 °F)
peak 30 min/day: -30 - 140 °C (-22 - 284 °F)

Ex-proof OPTION EX (EN-D)

Sensor assembly in ex-proof version accor. ATEX (EN-D)

Ex-proof OPTION EX (FM-D)

Sensor assembly in ex-proof version accor. FM (FM-D)

C4000 - C8000 Product Information

Visit our documentation page to view a complete list of translated documents.